FLUORAT®-02-PANORAMA Spectrofluorometer
|
|
FLUORAT®-02-PANORAMA spectrofluorometer is the most sophisticated instrument among the FLUORAT®-02 family. It is intended for scientific research, chromatographic analysis, and for operation as part of multimodule automatic measuring facilities. The spectral regions in the excitation and detection paths are set by built-in fast monochromators. The monochromators are controlled independently, so that the analyst can program any of their operation functions in the process of measuring, in particular, the spectra can be synchronously scanned.
A proprietary optical scheme provides the high sensitivity of the
instrument, especially in the UV spectral range, where the isolation
of a spectral interval using a light filter is difficult. The instrument
is controlled either from a built-in keypad or from an external
computer. The program-controlled computer interface implements the
«virtual device» operating mode. In this mode, the current status
of all the units and processes occurring in the instrument are displayed
on the screen.
Description
|
|
Extended capabilities:
- Analysis of liquid and solid samples by the luminescence spectra
- Chromatographic analysis of several components during single eluation
- Synchronous and two-dimensional scanning of the
spectra in both excitation and luminescence detection
paths
- Precise measurements of luminescent kinetic processes.
- Detecting and recording of luminescence and luminescence
excitation spectra, or synchronous luminescence scanning
spectra.
|
|
FLUORAT®-02 PANORAMA incorporates two monochromators in the luminescence
excitation and detection paths, which are arranged in a compact
two-level configuration, a Xenon flashlamp as a light source,
and a PMT as a detector of luminescence signals. To enhance the
measurement accuracy, a correction with respect to the reference
and transmission paths is used. The monochromators employ 900
- l/mm concave diffraction gratings with 100-mm radius of curvature.
The wavelength is set either from a keyboard or using a computer
control program.
| SPECIFICATIONS |
| Spectral range in the excitation channel, nm |
210 - 840 |
| Spectral range in the transmission channel, nm |
210 - 840 |
| Spectral range in the luminescence channel, nm |
210 - 690 (1) |
| Spectral bandwidth of monochromators, nm |
< 8 (15) (2) |
| Wavelength accuracy, nm |
< 3 |
| Signal-to-noise ratio (3) |
> 100 (200)(4) |
| Volume of the analyzed probe in a standard Q10 cuvette, mL |
3 |
| Power requirements |
110/220 Vac, 50/60Hz, 40 W |
| Dimensions, mm (inches) |
400x360x160 (16x14x6) |
| Weight, kg (lbs.) |
13 (29) |
|
(1) Upon customer's request.
(2) 8 nm is recommended for spectroscopic, and 15 nm for chromatoghrafic application.
(3) For Raman scattering of water at the excitation wavelength of 350 nm (emission at 400 nm), with the time response of 2 sec.
(4) Signal-to-noise ratio is more then 100 for the instruments with the bandwidth of 8 nm and 200 for the instruments with the bandwidht of 15 nm.
|
 |